Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Temperature range: -40 to 120 °C
JTM K07
Temperature fluctuation: ±0.3°C
Temperature gradient: ±2.5°C
Spatial temperature deviation: ±2.0°C
JTM K09 & K07
Temperature change rate (Rise): 2.5°C/min. (-24 to 104°C)
Temperature change rate (Drop): 0.8°C/min. (104 to -24°C)
Time to reach extreme temperatures (Rise): Within 100 min. (20 to 120°C)
Time to reach extreme temperatures (Drop): Within 100 min. (20 to -40°C)
Testing room floor space: 3.3 m²
External dimensions (W x D x H): 2365 x 1846 x 2330 mm
Internal dimensions (W x D x H): 1500 x 1500 x 1900 mm
Exterior/interior materials: Color steel plate (ivory) / stainless steel plate
Floor load capacity: 5.9 kN/m²
Door (width x height): 830 x 1,800 single opening, 1 location
Observation window (W x H): 190 x 320 door area, 1 location
Inner light (60 W capacity): 1 Qty.
Cable hole: ø50, 1 location (with rubber plug)
Defrosting method: Off-cycle defrosting (refrigerator operation stopped, room temperature 5 to 40℃), temperature-increase defrosting (room temperature -40/-10 to 5℃)
Power supply: Three-phase 200V 50/60Hz
Maximum load current: 45 A
ELB capacity: 75 A
Coolant
Water volume: 2400 L/h
Water pressure: 0.1 to 0.5 MPa
Water temperature: 18 to 32 °C
Piping dimensions inlet/outlet: Rc1 / Rc1