Differential Pressure Measurement

Instrument: 2kPa, 20kPa

Range: ±2000.0Pa / ±999.99Pa, ±20000Pa / ±9999.9Pa

Accuracy @ 20°C (including linearity and hysteresis): < 0.1% R + 0.03 Pa, < 0.1% R + 0.3 Pa

Control Stability (if internal pump fitted): <0.003% FS (with both ports sealed)

Resolution: 1:200000 (auto-range 5-digit display)

Temperature Coefficients:

Zero: < 0.02 % fsd per °C (eliminated by Automatic zero function)

Span: < 0.01 % per °C

Measurement inputs

Ranges:

3 x current inputs: 4-20 mA (0.001mA resolution)

1 x Voltage input: ± 20 VDC (0.1mV resolution)

Accuracy @ 20°C:

4 to 20 mA input: < 0.025% + 0.002 mA

±20 VDC Input: < 0.025% + 0.2 mV

PT100 input: Measurement range: 0 to 50 °C (0.01°C resolution) Accuracy <0.025% + 0.004Ω

Absolute Pressure: Range: 600 to 1600 mbar absolute. (0.1 mbar resolution) Accuracy < 0.5%FS

Electrical

Supply Voltage: 12 VDC ± 10% 2A (Universal mains power supply included), 2.1mm power socket

Measurement inputs

Test: 8-pin socket ( voltage + current)

External Pressure Sensor: 5-pin socket (4-20mA)

External Temperature Sensor: 5-pin socket (4-20mA)

External PT100 Sensor: 4-pin socket (2, 3 or 4-wire connection)

Communication Interfaces

RS232: 9-pin D plug

USB: Type A connector for connection to EUT or memory stick

USB: Type B connector for connection to PC

Interface

Digital I/O: 4 inputs/outputs (5V logic)

Analogue output: 0 to 10VDC

General

Enclosure: Aluminium enclosure with tilt feet

Display: Blue backlit graphic display.

Datalogger capacity: Storage of approximately 35,000 single display readings.

Dimensions: W 250mm x H 105mm x D 265mm (without handle)

Operating temperature: +5°C to +45°C

Storage Temperature: -10°C to +50°C

Battery life: > 8 hours of normal use

Weight: < 4.5 kg

Long Term Drift (span): < 0.25% reading per year

Overload: 20 times fsd without damage

Media Compatibility: Clean dry air or non-corrosive gas

Static pressure: Without internal pump: -1 bar to + 4 bar. With internal pump: -500mbar to + 500 mbar.

Pneumatic Connections

Test Ports: Push-on tube fittings for 6mm o/d, 4mm i/d flexible tube.

Absolute sensor port: M5 female thread.

Options:

Basic unit (Without pressure generator or battery)

Automatic pressure generator

Internal Absolute pressure sensor

Internal battery pack

Soft carry case