Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Type: Stationary sample type, cool/hot air switchable, two/three zone switchable
Temperature setting range:
Testing temperature: Low-temperature testing: −70 to 0℃, High-temperature testing: 60 to 200℃
Precooling temperature: −80℃ to testing chamber temperature low-temperature setting
Preheating temperature: Testing chamber temperature high-temperature setting to 210℃
Temperature fluctuation margin: ±0.5℃
Temperature rise time (high-temperature chamber): Room temperature to 200℃: Within 20 min.
Temperature drop time (low-temperature chamber): Room temperature to −75℃: Within 50 min.
Temperature cycle performance:
Temperature return time: Within 5 min.
Return time guaranteed cycles: 5 cycles (low-temperature start)
Testing conditions:
Sample Type: Plastic molded IC
Sample Weight: 6.5 kg (IC 4.5 kg + dual-stage jig 2.0 kg)
Testing chamber temperature / Testing time (Low): -65℃/30 min.
Testing chamber temperature / Testing time (Room): Atmosphere temperature/5 min.
Testing chamber temperature / Testing time (High): 150℃/30 min.
Temperature measurement position: Upwind side
Continuous operation cycles: 100 cycles (low-temperature start)
Testing chamber volume: 74 L
Permissible sample weight inside testing chamber: Max. 30 kg (Bottom 20 kg or less, sample shelf 5 kg or less [per shelf])
Testing chamber dimensions (W × D × H): 410 × 390 × 460 mm
Product dimensions (W × D × H): 1,380 × 1,400 × 1,980 mm
System component materials:
Exterior: Stainless steel plate (SUS430, hairline finish)
Interior: Stainless steel plate (SUS304, 2B polished)
Cooling method: Dual refrigeration using mechanical compressor (water-cooling)
Cooling system refrigerant/output:
High-temperature side: R404A/4.5 kW (inverter control)
Low-temperature side: R23/3.7 kW (inverter control)
Safeguards: Earth leakage circuit breaker, electric motor overload safeguard, high-voltage shutdown system, temperature fuse, operation with door open prevention switch, excess temperature increase prevention system, and protection fuse
Operational noise: 65 dB(A)
Accessories: Cable hole soft silicon plug × 1, sample shelf × 2, shelf support × 4, operation circuit protection fuse × 2, coolant piping Y strainer × 2 sets, and instruction manual × 1
Power supply: Three-phase 200V 50/60Hz, Three-phase 220V 60Hz, Three-phase 380V 50Hz
Maximum load current / Power supply capacity: 78 A / 27 kVA
Dry air (Pressure): 0.4 to 0.7 MPa
Dry air (Consumption): Approx. 10.2 L(ANR) per cycle
Coolant:
Water volume: 2,750 L/h (coolant inlet temperature 32℃)
Water pressure: 0.1 to 0.5 MPa
System side piping diameter: Rc 1 1/4
Drain piping: Rc 1/2
Product weight: 950 kg