Digital MultiMeter Calibration Procedure

04/06/2021 10:26:04

**1.Range of application** \- DC Voltage 1000V \- AC Voltage 1000V, Frequency 100kHz \- DC Current 30A \- AC Current 30A, Frequency 10kHz \- Resistor 100

1.Range of application

- DC Voltage 1000V

- AC Voltage 1000V, Frequency 100kHz

- DC Current 30A

- AC Current 30A, Frequency 10kHz

- Resistor 1000MΩ

2. Calibrating techniques

- Outside check

- Technical check

- Measurement check

3. Calibrating instruments

3.1 Standard measuring instrument (s): Multifunction Calibrator (MFC)

Meatest M142 Multifunction Calibrator 15 ppm

- DC Voltage: 0µV - 1000V.

- AC Voltage: 0,1mV - 1000V, Frequency: 20Hz - 100kHz.

- DC Current: 0µA - 30A.

- AC Current: 1µA - 30A, Frequency: 20Hz - 10kHz.

- Resistor: 0 Ω - 1GΩ.

3.2 Auxiliary Instrument(s)

3.2.1 Testo Hygro-Thermometer 608-H1

Testo 608-H1 - Thermal Hygrometer (Part Number 0560 6081) | TEquipment

- Range of Measurement: (0 ÷ 50) °C; Accuracy: ±0,5 °C

- Range of Measurement: (10 ÷ 95) %RH; Accuracy: ±3 %RH

3.2.2 Dedicated Measuring Probe

- Probes with small electromotive force.

4. Conditions for Calibration

- Temperature: (23 ± 5) °C.

- Humidity: no higher than 80% RH.

5. Prepare for Calibration

- Make sure the power supply to the DMM meets the requirements specified in the catalog, the fuses, the protective circuit are still working properly.

- The DMM should be placed in the calibration environment for at least 8 hours.

- MFC and DMM must be turned on and dried for at least 30 minutes prior to calibration.

6. Begin the calibration

a. Outside Check

b. Technical Check

c. Measurement Check

    1. General requirements

    2. Begin the DMM Calibration

        a. DC Voltage

        b. AC Voltage

        c. DC Current

        d. AC Current

        e. Resistor

7. Measurement Uncertainty Evaluation

8. General handling

a. The calibrated DMM is stamped, certified in paper along with calibration results, and measurement uncertainty.

b. The recommended calibration cycle of DMM is 12 months.


For calibration of DMM, click here.

For further information about Digital MCalibration service, please contact

Equipment Verification - Calibration - Repair Service Center - EMIN Vietnam JSC 

Address: No 203 Nguyen Ngoc Vu Street, Trung Hoa Ward, Cau Giay District, Ha Noi City

Email: services@emin.vn - Hotline: 0968.060.220


Related News

Connecting and configuring the optical amplifier optimized for the line
09/05/2026 10:39:06

In many automated workshops, control systems occasionally experience issues where output signals freeze or become unstable, or fail to accurately detect minuscule components passing through narrow gaps on the conveyor belt, causing the machining process to halt abruptly without an apparent root cause.

The AI ​​Era: GPT-6 Astra Launches. Why Is OpenAI Pushing AI into a New Era?
09/04/2026 10:48:38

GPT-6 Astra has just launched and immediately became one of the most searched AI terms in the U.S., garnering over 200,000 searches in just 24 hours. The significance lies not in the name "GPT-6," but in a much broader shift: AI is moving from merely providing answers to directly performing tasks on the computer.

Chroma Integrates MXO Oscilloscope Hardware into the ATS 8000 Automated Test Platform
09/04/2026 09:32:17

Chroma has integrated Rohde & Schwarz MXO oscilloscope hardware into its ATS 8000 automated test platform. The integration expands waveform measurement and analysis capabilities for testing power conversion devices, meeting the increasingly demanding requirements of high-power supplies and power systems used in data centers.

Stay Updated with Offers

Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.

By subscribing, you agree to our Terms of Service and Privacy Policy.

Quick Support

Direct access to our certified experts