Gastec 1D Passive Dosl Tubes Carbon monoxide CO (1.04 ~ 2000 ppm)
ManufacturerGastec(View more products from this manufacturer)
Model1D
Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available
Indicator: 50~1000ppm Mendicator: 50~1000ppmMendicator: 50~1000ppm
Measuring Range: 1.04 to 2000 ppm
Sampling Hours: 0.5 to 48 hours
Colour Change: Pale Yellow → Brown
Reaction Principle: CO + Na2Pd(SO3)2 → Pd + CO2 + SO2 + Na2SO3
Coefficient of Variation: 10% (for 50 to 1000 ppm-hr)
Nr. of tests: 10
Performance
| Measuring Range | 1.04 to 2000 ppm |
| Sampling Hours | 0.5 to 48 hours |
| Detecting Limit | 2 ppm (10 hours) |
| Colour Change | Pale Yellow → Brown |
| Reaction Principle | CO + Na2Pd(SO3)2 → Pd + CO2 + SO2 + Na2SO3 |
| Coefficient of Variation | 10% (for 50 to 1000 ppm-hr) |
| Shelf Life | 2 Years |
| Corrections for temperature & humidity | Unnecessary |
| Store the tubes at cool and dark place. | |
Possible coexisting substances and their interferences (NOTE)
| Substance | Concentration | Interference | Change colour by itself to |
| Nitrogen dioxide | ≧ 5 ppm | No | |
| Chlorine | No | No | |
| Hydrogen sulphide | + | Reddish brown |
| Calibration gas generation | High pressure gas cylinder method |
| TLV-TWA | TLV-STEL | Explosive range |
| 25 ppm | - | 12.5 to 74% |
Stay Updated with Offers
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Quick Support
Direct access to our certified experts