OPERATING AMBIENT CONDITIONS
Temperature : 18 °C to 28 °C
Humidity: Up to 80 % RH
CONTROLLED RANGE
With dry nitrogen option
With ultrasonic option
Temperature 0 °C to 60 °C
Humidity @18 °C : 1 % to 95 %RH
Humidity @30 °C : 1 % to 80 % RH
Humidity @23 °C : 1 % to 99 %RH
Humidity @40 °C : 1 % to 60 % RH
Humidity @28 °C: 1 % to 95 % RH
Humidity @50 °C : 1 % to 50 % RH
CALIBRATION SYSTEM ACCURACY
Temperature: ± 0.1 °C Typically ± 0.05 °C Based on Probe Accuracy
Humidity: ± 0.50 % RH Or Better Typically ± 0.30 % Based on Probe Accuracy
Controller type : PID Controller
UNIFORMITY
Temperature : 0.10 °C
Humidity: 0.30 % RH @ 23 °C
STABILITY
Temperature: 0.05 °C
Humidity : 0.10 % RH @ 23 °C
RESOLUTION : DISPLAY
Temperature : 0.01 °C
Humidity : 0.01 % RH
Dew Point: 0.01 °C (Calculated)
RAMP / SOAK RATE OF CHANGE
Depends on ambient and desiccant conditions
Temperature
High → Low 0.50 °C / minute *Typical (Based on Desiccant)
Low → High 1.00 °C / minute *Typical (Based on Desiccant)
Humidity
High → Low 1.00 % RH / minute *Typical (Based on Desiccant)
Low → High 2.50 % RH / minute *Typical (Based on Desiccant)