Work voltage: AC220V ± 10%, 50Hz
Test range: 5-93 ° C (the instrument can reduce the temperature of the sample to -5 ° C)
Heating rate: flash point <60 ° C (0.9 ~ 1.1 ° C/min); flash point ≥60 ℃ (2.7 ~ 3.3 ℃/min)
Detection interval: flash point <60 ℃ (0.5 ℃); flash point ≥60 ℃ (1 ℃)
Device maximum power: not greater than 300W (work when each component is different)
Heating power: 150W; cooling power: 160W; ignition stick power: 120W
Operation interface: 7 -inch high -resolution LCD touch screen, resolution 800*480.
Temperature display accuracy: 0.1 ℃
Per pressure detection: Digital pressure sensing chip, error ± 0.3kpa @25 ℃.
Refrigeration parts: semiconductor refrigeration film, with water flow switch protection function.
Refrigeration and heat exchange method: Cooling water cooling, randomly equipped with a submersible pump
Printer: Embedded thermist micro printer
Historical records: It can store 256 groups of historical results values
Hardware test: with switching cup lid test, ignition mechanism test, stirring test, ignition stick test, heating test, refrigeration test