Weight: 900 g
Compliance: CE
Fill Fluid: NEOBEE M5 food grade
Weight Note: 900 g (2 lb) Typical depending upon configuration
Battery life: 9 mo
Cable Length: 2.7 m
Display Type: 4 Digit LCD
Span Setting: ± 0.5% FS typical ± 1% maximum (For ranges ≤2.5 psi ± 1% Typical ± 2% maximum)
Accuracy Note: ± 0.08% Best Straight Line Typ. (Linearity, Hysteresis and Repeatability combined)
Output Signal: Current, Voltage
Pressure Type: Compound Gauge
Pressure, Max: 5 psi
Pressure, Min: -5 psi
Wireless Band: None
Accuracy Class: < 0.1%
Burst Pressure: 6x Full Scale
Data Interface: USB
Pressure Cycles: 250000
Thermal Effects: Thermal Effect (Over Compensated Range): Zero and Span shift over Compensated Range; Ranges: ≤2.5 psi: Zero: 1.00% Span: 0.70% ;5 psi: Zero: 0.80% Span: 0.50% ; > 5 psi: Zero: 0.50% Span: 0.50% ; 26B & 26HB Barometric Range: Zero: 0.90% Span: 0.50%
Calibration Note: 5-Pt NIST traceable Calibration with zero and span values, Calibrated in horizontal direction (Compound Gage models calibrated in positive direction only)
Number of Digits: 4
Display Type Note: 4-digit LCD with user selectable backlighting, (9999 count) 25 mm (1') character height.
Electrical Output: 4 to 20 mA
Enclosure Ratings: IP65, NEMA Type 4X, Weatherproof
Materials, Wetted: 316 Stainless Steel
Pressure Accuracy: ± 0.08 %
Software Settings: Software Settings (via USB) Update Rate: 4 options, 10, 5, 2, 1 times per second Units: psi, inHg, inH2O, bar, mbar, hPa units; must be specified at time of order Lock: Allows the lock-out of front panel functions Alarms: User selectable “High” and “Low” alarm limits (open collector) Backlight: On/off, 10 sec, 30 sec, 1 min, 5 min Wireless Transmitter (Optional): Channel number, transmission rate, alarms, sensor offset, chart recording, data logging Calibrate: Zero and Span Overrange Indication: “_ _ _ _”
Supply Power Note: Battery Note:One 3.6V 8.4 Ah capacity (C cell) Included (Two included with Wireless Option)
Supply Power Type: Battery
Enclosure Material: 316 Stainless Steel
Zero/Balance Error: ± 0.5% FS typical ± 1% maximum (For ranges ≤2.5 psi ± 1% Typical ± 2% maximum)
Connection Location: Lower
Data Interface Note: USB connection for set-up, cable included
Number of Batteries: 1
Sample/Display Rate: 0.25 to 10 sec
Settable Parameters: Backlight, Min, Max, Zero, Clear
Thermal Span Effect: 0.005
Thermal Zero Effect: 0.008
Battery Size and Type: 3.6V Li-ion
Sensor Dimension Note: 72 to 89 mm L (2.8 to 3.5") below display
Display Segment Height: 25 mm
Electrical Output Note: User selectable 0 to 5 Vdc, 0 to 10 Vdc of 4 to 20 mA (external 24 Vdc power required)
Process Connection Note: 11 ⁄2 and 2" Tri-GripTM Fitting (compatible with Tri-Clamp®): 600 psi using suitable clamp (600 psi on 11 ⁄2" Tri-Clamp® only)
Process Connection Size: 2 in
Process Connection Type: Tri-Clamp
Process Temperature, Max: 60 °C
Process Temperature, Min: 4.4 °C
Sample Display Rate Note: User selectable from 0.25 to 10 sec. (preset to 1/sec)
Storage Temperature, Max: 82 °C
Storage Temperature, Min: -40 °C
Operating Temperature, Max: 66 °C
Operating Temperature, Min: -18 °C
Wireless Option Specifications: Transmit Sample Rate: User programmable from 1 sample/2 min to 1 sample/2 sec Radio Frequency (RF) Transceiver Carrier: ISM 2.4 GHz RF Output: 10 dBm (10 mW) Range of RF Link: Outdoors (Line of Sight): Up to 120 m (400') Indoors/Urban: Up to 40 m (130') Included Software: Requires Windows® 7 (32 bit); XP, Vista Internal Battery: Two 3.6V lithium 8.4 Ah capacity (C-cell) included Data Transmission to Host: Pressure reading, ambient temperature reading, RF transmit strength and battery level
Isolation between Case and Output Terminations, Min: 100MΩ @ 50 Vdc